FACTORS AFFECTING SYSTEM SELECTION (INSTRUMENTATION)

LINEARITY

The linearity of the transducer is an expression of the extent to which the actual measured curve of a sensor departs from the ideal curve. Figure (3) shows a somewhat exaggerated relationship between the ideal, or least squares fit, line and the actual measured or calibration line (Note in most cases, the static curve is used to determine linearity, and this may deviate somewhat from a dynamic linearity) Linearity is often specified in terms of percentage of nonlinearity, which is defined as:

where

Nonlinearity (%) is the percentage of nonlinearity
Din(max) is the maximum input deviation
INf.s. is the maximum, full-scale input

The static nonlinearity defined by Equation 6-1 is often subject to environmental factors, including temperature, vibration, acoustic noise level, and humidity. It is important to know under what conditions the specification is valid and departures from those conditions may not yield linear changes of linearity.



Figure 3. Ideal versus measured curve showing linearity error. Source: J J Carr, Sensors and Circuits Prentice Hall

ACCURACY

The accuracy of the sensor is the maximum difference that will exist between the actual value (which must be measured by a primary or good secondary standard) and the indicated value at the output of the sensor. Again, the accuracy can be expressed either as a percentage of full scale or in absolute terms.

PRECISION

The concept of precision refers to the degree of reproducibility of a measurement. In other words, if exactly the same value were measured a number of times, an ideal sensor would output exactly the same value every time. But real sensors output a range of values distributed in some manner relative to the actual correct value. For example, suppose a pressure of exactly 150 mm Hg is applied to a sensor. Even if the applied pressure never changes, the output values from the sensor will vary considerably. Some subtle problems arise in the matter of precision when the true value and the sensor's mean value are not within a certain distance of each other
(e.g., the 1-s range of the normal distribution curve).


SENSITIVITY


The sensitivity of the sensor is defined as the slope of the output characteristic curve (DY/DX in Figure 1) or, more generally, the minimum input of physical parameter that will create a detectable output change. In some sensors, the sensitivity is defined as the input parameter change required to produce a standardized output change. In others, it is defined as an output voltage change for a given change in input parameter. For example, a typical blood pressure transducer may have a sensitivity rating of
10 mV/V/mm Hg; that is, there will be a 10-mV output voltage for each volt of excitation potential and each mm Hg of applied pressure.

Sensitivity Error
The sensitivity error (shown as a dotted curve in Figure 1) is a departure from the ideal slope of the characteristic curve. For example, the pressure transducer discussed above may have an actual sensitivity of 7.8 mV/V/mm Hg instead of 10 mV/V/mm Hg.





Figure 1. Ideal curve and sensitivity error. Source: J.J. Carr, Sensors and Circuits Prentice Hall.


HYSTERESIS



A transducer should be capable of following the changes of the input parameter regardless of which direction the change is made; hysteresis is the measure of this property. Figure 4 shows a typical hysteresis curve. Note that it matters from which direction the change is made. Approaching a fixed input value (point B in Figure 4) from a higher value (point P) will result in a different indication than approaching the same value from a lesser value (point Q or zero). Note that input value B can be represented by F(X)1, F(X)2, or F(X)3 depending on the immediate previous value—clearly an error due to hysteresis.



Figure 4. Hysteresis curve. Source: J.J. Carr, Sensors and Circuits Prentice Hall.

RANGE

The range of the sensor is the maximum and minimum values of applied parameter that can be measured. For example, a given pressure sensor may have a range of -400 to +400 mm Hg. Alternatively, the positive and negative ranges often are unequal. For example, a certain medical blood pressure transducer is specified to have a minimum (vacuum) limit of -50 mm Hg (Ymin in Figure 1) and a maximum (pressure) limit of +450 mm Hg (Ymax in Figure 1). This specification is common, incidentally, and is one reason doctors and nurses sometimes destroy blood pressure sensors when attempting to draw blood through an arterial line without being mindful of the position of the fluid stopcocks in the system. A small syringe can exert a tremendous vacuum on a closed system.


Dynamic Range
The dynamic range is the total range of the sensor from minimum to maximum. That is, in terms of Figure 1, Rdyn = Ymax - l -Yminl.

DYNAMIC LINEARITY


The dynamic linearity of the sensor is a measure of its ability to follow rapid changes in the input parameter. Amplitude distortion characteristics, phase distortion characteristics, and response time are important in determining dynamic linearity. Given a system of low hysteresis (always desirable), the amplitude response is represented by:
F(X) = aX + bX2 + cX3
+ dX4 + ••• + K (6-2)

In Equation 6-2, the term F(X) is the output signal, while the X terms represent the input parameter and its harmonics, and K is an offset constant (if any). The harmonics become especially important when the error harmonics generated by the sensor action fall into the same frequency bands as the natural harmonics produced by the dynamic action of the input parameter. All continuous waveforms are represented by a Fourier series of a fundamental sinewave and its harmonics. In any nonsinusoidal waveform (including time-varying changes of a physical parameter). Harmonics present will be that can be affected by the action of the sensor.




Figure 6. Output versus input signal curves showing (a) quadratic error; (b) cubic error. Source: J.J. Carr, Sensors and Circuits Prentice Hall.

The nature of the nonlinearity of the calibration curve (Figure 6) tell something about which harmonics are present. In Figure 6a, the calibration curve (shown as a dotted line) is asymmetrical, so only odd harmonic terms exist. Assuming a form for the ideal curve of F(x) = mx + K, Equation 6-2 becomes for the symmetrical case:
F(X) = aX + bX2 + cX4 + ••• + K (6-3)

In the other type of calibration curve (Figure 6b), the indicated values are symmetrical about the ideal mx + K curve. In this case, F(X) = -F(-X), and the form of Equation 6-2 is:
F(X) = aX + bX3 + cX5 + ••• + K (6-4)

Now we will take a look at some of the tactics and signals processing criteria that can be adapted to biomedical applications to improve the nature of the data collected from the sensor.


RESOLUTION

This specification is the smallest detectable incremental change of input parameter that can be detected in the output signal. Resolution can be expressed either as a proportion of the reading (or the full-scale reading) or in absolute terms.

OFFSET

The offset error of a transducer is defined as the output that will exist when it should be zero or, alternatively, the difference between the actual output value and the specified output value under some particular set of conditions. An example of the first situation in terms of Figure 1 would exist if the characteristic curve had the same sensitivity slope as the ideal but crossed the Y-axis (output) at b instead of zero. An example of the other form of offset is seen in the characteristic curve of a pH electrode shown in Figure 2. The ideal curve will exist only at one temperature (usually 25°C), while the actual curve will be between the minimum temperature and maximum temperature limits depending on the temperature of the sample and electrode.



Figure 2. Typical pH electrode characteristic curve showing temperature sensitivity. Source: J.J. Carr, Sensors and Circuits Prentice Hall.





RESPONSE TIME


Sensors do not change output state immediately when an input parameter change occurs. Rather, it will change to the new state over a period of time, called the response time (Tr in Figure 5). The response time can be defined as the time required for a sensor output to change from its previous state to a final settled value within a tolerance band of the correct new value. This concept is somewhat different from the notion of the time constant (T) of the system. This term can be defined in a manner similar to that for a capacitor charging through a resistance and is usually less than the response time.

The curves in Figure 5 show two types of response time. In Figure 5a the curve represents the response time following an abrupt positive going step-function change of the input parameter. The form shown in Figure 5b is a decay time (Td to distinguish from Tr, for they are not always the same) in response to a negative going step-function change of the input parameter.




Figure 5. (a) Rise-time definition; (b) fall-time definition. Source: J.J. Carr, Sensors and Circuits Prentice Hall. 

Comments

Popular posts from this blog

CCNA3 v4.0 Final Exam Answers Updated 2013 100%

1. Refer to the exhibit. A network administrator needs to add IP phones to the network. To which devices should the IP phones connect? AS1 and AS2 DS1 and DS2 DS1, DS2, and CS1 AS1, AS2, DS1, and DS2 2. Which CLI mode allows users to access all device commands, such as those used for configuration, management, and troubleshooting? user EXEC mode privileged EXEC mode global configuration mode interface configuration mode 3. Refer to the exhibit. Which Spanning Tree Protocol version is in effect? Per VLAN Spanning Tree (PVST) Per VLAN Spanning Tree + (PVST+) Common Spanning Tree (CST) Rapid Spanning Tree Protocol (RSTP) Multiple Spanning Tree Protocol (MSTP) 4. Refer to the exhibit. A network administrator has segmented the network into two VLANs and configured Router1 for inter-VLAN routing. A test of the network, however, shows that hosts on each VLAN can only access local resources and not resources on the other VLAN. What is the most likely cause of this pr

CCNA3 v3.0 Final Exam Answers Updated 2013 100%

1. Refer to the exhibit. The Layer 2 switching design that is shown has been implemented in a campus environment that is using Spanning Tree Protocol. All inter-switch links that are shown are trunks. Whenever an inter-switch link fails, the network takes nearly a minute to completely converge. How can the convergence time be reduced? Increase the capacity of the distribution and core trunk links to 10 Gb/s. Add a trunk link that directly connects D1 and D2. Use Layer 3 switching on the core switch. Implement Rapid Spanning Tree Protocol 2. Refer to the exhibit. R1 is configured for traditional inter-VLAN routing. R1 can ping computer 3 but cannot ping computer 1. What is a possible cause for this failure? S1 port Fa0/11 is in the wrong VLAN. R1 does not have an active routing protocol. The IP address of computer 1 is in the wrong logical network. Router interface Fa0/0 has the wrong trunk encapsulation type configured. 3. Refer to the exhibit. What will allow a h

CCNA 2 EROUTING version 4 2013 FINAL EXAM ANSWERS 100%

CCNA 2 Final Exam 2013 v 4   1. Refer to the exhibit. A network engineer removes a new router from the shipping container and powers on the router to ensure it passes POST. Which port would the engineer use to perform the initial configuration? AUX console FE0/0 FE0/1 2. What is the purpose of the TTL field within an IP packet header? clears an unreachable route from the routing table after the invalid timer expires prevents regular update messages from inappropriately reinstating a route that may have gone bad removes an unreachable route from the routing table after the flush timer expires limits the period of time or number of hops a packet can traverse through the network before it should be discarded used to mark the route as unreachable in a routing update that is sent to other routers 3. When would the network administrator use the ip bandwidth-percent eigrp as-number percent command? when there is a low bandwidth connection when the connec